Comparison of noninvasive continuous arterial blood pressure measured by NICAP with arterial line in elderly patients

A Correction to this article was published on 28 February 2022

This article has been updated

Abstract

Background

Non-Invasive Continuous Arterial Pressure system (NICAP) allows continuous monitoring, timely detection of hypotension, and avoiding risks from invasive procedures. A previous study showed good comparability of NICAP with arterial line in people with no evidence of cardiovascular disease. Therefore, the goal of this study was to investigate whether NICAP could be accurately applied to elderly patients.

Methods

In this single-centered observational study, forty-one patients above 65 undergoing elective surgeries requiring artery catheterizations were enrolled from July 17, 2020, to June 25, 2021. Radial artery cannulation and NICAP monitoring were started before anesthesia. Blood pressure during the anesthesia induction and the whole surgery, trend of blood pressure changes, time needed for establishing continuous monitoring, and complications were recorded.

Results

A total of 6751 valid pairs of blood pressure measurements were analyzed. In the Bland-Altman analysis, the arithmetic means for systolic, diastolic, and mean arterial pressure were 2.2, 3.3, and 2.8 mmHg, respectively. NICAP and arterial line correlation coefficients for systolic, diastolic, and mean arterial pressure were 0.51, 0.40, and 0.47, respectively. In the trending analysis, the polar concordance rates at 30 degrees were 70.9% for systolic, 67.7% for diastolic, and 69.3% for mean arterial blood pressure. During the anesthesia induction, the arithmetic means for systolic, diastolic, and mean arterial pressure in the Bland-Altman analysis were 1.7, -0.2, and 0.5 mmHg, respectively. NICAP and arterial line correlation coefficients for systolic, diastolic, and mean arterial pressure were 0.72, 0.58 and 0.69, respectively. No severe complications occurred.

Conclusions

NICAP has a poor correlation with the arterial line in elderly patients for the whole surgery or during anesthesia induction. Moreover, it showed poor comparability in the detection of blood pressure change trends with arterial lines. Our findings suggest that NICAP might not be sufficiently accurate to be applied clinically in elderly patients with comorbidities. More accurate calibration and iteration are needed.

Peer Review reports

Background

Intraoperative hypotension, whether absolute decrease in mean arterial pressure or hypotensive episodes, is strongly associated with poor postoperative outcomes especially for elderly patients whose vessel elasticity and cardiopulmonary function reserve are significantly reduced due to common comorbidities, such as hypertension and diabetes [1,2,3,4,5,6,7,8,9]. Early detection of hypotension and individualized maintenance of blood pressure during surgery could reduce risks of postoperative complications and improve the prognosis for elderly patients [1, 9,10,11,12].

Current monitoring recommendation of blood pressure taken every 5 min could neglect some hypotension episodes while arterial catheterization as the gold standard could cause severe complications [13, 14]. Non-invasive continuous blood pressure monitoring could be an excellent substitute. Several non-invasive continuous blood pressure monitoring devices with various monitoring mechanisms have been developed such as Nexfin, T-line tensimeter, and Continuous Non-invasive Arterial Pressure (CNAP). Those devices need calibration and correction from cuff blood pressure to form accountable blood pressure readings and took a relatively long time to establish. Non-Invasive Continuous Arterial Pressure system (NICAP) is another monitoring device using two near-infrared light sensors to detect the arterial pulse waveforms of the ear and toe to calculate the pulse wave velocity and formulate arterial blood pressure [15,16,17]. Its application is simple, quick, and could avoid complications associated with invasive procedures. In addition, due to its mechanism, intermittent calibration from cuff blood pressure is not needed. In a previous validation study, NICAP showed good comparability with invasive blood pressure monitoring and met The Association for the Advancement of Medical Instrumentation (AAMI) standard in people with no evidence of cardiovascular disease [18]. However, the application of NICAP in elderly patients who may have several comorbidities has never been validated before.

Thus, we conducted this observational study to investigate whether NICAP could accurately reflect arterial line blood pressure in elderly patients during surgery. We hypothesized that NICAP was accurate compared with arterial line blood pressure in elderly patients.

Methods

Study design

The Ethics Committee of Clinical and Biomedical Trials, West China Hospital of Sichuan University (No. 2019-136). Written informed consent was obtained from all the participants. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). Forty-one patients were enrolled from July 17, 2020, and June 25, 2021. Inclusion criteria were: (1) over the age of 65, (2) undergoing elective surgery in a supine position, (3) requiring arterial catheterization. Patients were excluded if they had: (1) advanced dysfunction of peripheral perfusion, (2) arteriovenous shunts for hemodialysis, (3) vascular surgery of lower extremities, (4) peripheral arterial disease, and (5) anatomical abnormalities or injuries at the NICAP sensor site.

NICAP monitoring technique

NICAP technology (Zhejiang Union Medical Equipment Co., Ltd.) uses two near-infrared light sensors to detect the arterial pulse waveforms of the ear and toe for calculating the pulse wave velocity (PWV). The system uses instantaneous pattern changes of the pulse waveform to calibrate the PWV. A mathematical model algorithm was established between PWV and blood pressure based on the wave motion equation, elastic mechanics, and hemodynamics [15,16,17].

Blood pressure monitoring process

Before induction of anesthesia, radial arterial cannulation was performed for all the patients using 20G SURFLO® (TERUMO CHINA HOLDING CO., LTD.) under local anesthesia (2% lidocaine) and arterial blood pressure was monitored (Single Channel Transducer Kit for IBP Monitoring, SCW MEDICATH LTD, Shenzhen, China). Then the NICAP sensors were applied to the patients’ ears and toes according to the manufacturer’s recommendations (Fig. 1). The time needed for cannulation and NICAP monitoring were recorded, respectively. Recordings of systolic (SBP), diastolic (DBP), and mean blood pressure (MAP) from both methods were started at the induction of anesthesia and ended after the patients were transferred to post-anesthesia care unit (PACU) for 1-minute intervals in the anesthesia information system. During induction, blood pressure was recorded every 30 s for 5 min manually. Complications of both methods, including hematoma, hemorrhage, nerve injury, and pressure injury to the skin, were documented.

Fig. 1
figure 1

 A and B Application of the ear and toe clamps. C Monitor Screen of NICAP showing blood pressure, heart rate, and the pulse waveforms of ear and toe. (Figures cited from the NICAP user manual of Zhejiang Union Medical Equipment Co., Ltd. Witten consent of usage and adaptation was obtained from the company)

Anesthesia procedure

Monitoring of electrocardiography, pulse oximetry, capnography, and body temperature was started when patients arrived at the operating room (OR). After radial artery cannulation and calibration of arterial blood pressure, anesthesia was induced intravenously with midazolam 0.03 mg/kg, sufentanil 0.3ug/kg, propofol 1-2 mg/kg, and cis-atracurium 0.2 mg/kg. After intubation, anesthesia was maintained with 2% sevoflurane or 6% desflurane and remifentanil 0.05-0.2ug/kg/min. All patients were ventilated using a lung-protective strategy including tidal volume of 6-8ml/kg, positive end-expiratory pressure (PEEP) 6-8mmHg, and a frequency to maintain end-tidal carbon dioxide (EtCO2) between 35-40mmHg.

The primary outcome of this study was to assess the accuracy of NICAP compared with arterial blood pressure for elderly patients during the whole surgery (excluding the anesthesia induction). Secondary outcomes include blood pressure recorded from the two methods during anesthesia induction, the trend of blood pressure changes of the two methods, the time needed for establishing continuous monitoring for both methods, and complications.

Statistical analysis

There is no consensus on how many subjects and measurements should be analyzed when performing a repeated measure. Patients enrolled in different studies varied [19,20,21,22,23,24,25]. According to the AAMI protocol, validation of new monitoring tools for a specific population requires at least 35 patients. In this study, 41 patients were enrolled. A total of 6751 pairs of data were analyzed. All the data were blinded from the analyzer for statistical analysis. Data were expressed as the mean with standard deviations (SD) or median with the 25th to 75th percentiles (IQR). Categorical variables are shown as numbers (%). Data were tested for normal distribution with the Q-Q plots [26]. Blood pressure differences between methods were analyzed using a dependent t-test for paired measurements. Bland–Altman analysis for multiple measurements per subject was used to assess the bias (mean difference) and precision (SD of the bias) between SBP, DBP, and MAP recorded by the two methods. In addition, correlations between ABP and NICAP were determined by linear regression for the whole repeated paired measurements from all patients. Finally, the polar plots were drawn using SigmaPlot (Systat Software Inc, San Jose, CA) to analyze the two methods’ trending abilities of changes in SBP, DBP, and MAP [27,28,29]. Concordance rate above 92% was determined to be good, while below 90% was considered to be poor [27]. The Bland-Altman plots and scatter plots were drawn using MedCalc® Statistical Software version 19.7.2 (MedCalc Software Ltd, Ostend, Belgium; https://www.medcalc.org; 2021). Statistical analyses were performed using SPSS (IBM Corp. Released 2019. IBM SPSS Statistics for Macintosh, Version 26.0. Armonk, NY: IBM Corp). Two-sided p Values < 0.05 were considered statistically significant.

Results

A total of 60 potentially eligible patients were screened due to the surgical planning list. Forty-four consented, nine refused to participate, six failed to meet the inclusion criteria, and 1 was canceled due to poorly controlled hypertension. All patients received right radial artery cannulation after Allen’s test was normal. The NICAP waveform recording was determined to be acceptable in 44 patients. NICAP data were missing in 2 patients due to incorrect connection of the device. Radial artery canulation was failed in 1 patient. Finally, data from 41 patients were analyzed. A summary of the baseline characteristics of the patients is presented in Table 1. Total anesthesia time (mean ± SD) was 230.41 ± 98.66 min, and the time of surgery (mean ± SD) was 167.83 ± 86.21 min. Although we obtained 9487 pairs of blood pressure data during the study, 2736 pairs were excluded from the analysis, in which 1523 were due to arterial line flushing artifacts, 385 were due to arterial blood pressure data missing, and 828 due to NICAP data missing in the anesthesia information system. Therefore, 6751 valid pairs of simultaneous arterial (ART) and NICAP measurements were analyzed. The median blood pressure data pairs per patient was 123 (95-232). A total of 410 pairs of blood pressure measurements were obtained during anesthesia induction. 130 pairs from 7 patients were excluded from the analysis due to artifacts or incorrect connection. Thus, a total of 280 valid pairs of data from 34 patients were analyzed. The median blood pressure data pairs per patient during induction was 9 (7.75-10). The blood pressure results of the whole surgery and during anesthesia induction are shown in Tables 2 and 3. The time needed to establish monitoring for blood pressure was significantly shorter for NICAP compared with arterial catheterization. [ART vs. NICAP (min): 9 (7-13) vs. 2 (2-3), p < 0.001] There were no major complications that occurred either from arterial line or NICAP.

Table 1 Baseline characteristics of the patients
Table 2 Blood pressure measurements from all patients
Table 3 Blood pressure measurements during anesthesia induction from 34 patients

The results of the Bland-Altman plot for multiple measurements per subject between NICAP and ART are shown in Fig. 2. Bias and precision were 2.19 and 18.86 mmHg in SBP, 3.27 and 13.42 mmHg in DBP, 2.78 and 14.17 mmHg in MAP. Upper and lower limit of agreements (LOAs) were 38.74 and -34.36 mmHg in SBP, 30.84 and -24.31 mmHg in DBP, and 31.46 and -25.90 mmHg in MAP. Although bias fulfilled the AAMI criteria below 5 mmHg, precisions all exceeded the AAMI standard of 8 mmHg. The scatter plots for SBP, DBP, and MAP are shown in Fig. 3. The correlation coefficients were 0.51, 0.40, and 0.47 for SBP, DAP, and MAP, respectively. SBP, DBP, and MAP from NICAP were not strongly correlated with ART. The overall NICAP measurement for elderly patients was lower than ART regardless of SBP, DBP, or MAP. In the polar plot trending analysis, the polar concordance rate at 30 degrees was 70.9% for SBP, 67.7% for DBP, and 69.3% for MAP (Fig. 4). During anesthesia induction, the arithmetic means for systolic, diastolic, and mean arterial pressure in the Bland-Altman analysis were 1.7, -0.2, and -0.5 mmHg, respectively. Bias and precision were 1.72 and 18.31 mmHg in SBP, -0.17 and 12.59 mmHg in DBP, 0.49 and 12.34 mmHg in MAP. Upper and lower LOAs were 37.8 and -34.3 mmHg in SBP, 25.7 and -26.1 mmHg in DBP, and 25.5 and -24.5 mmHg in MAP (Fig. 5). NICAP and arterial line correlation coefficients for systolic, diastolic, and mean arterial pressure were 0.72, 0.58, and 0.69, respectively (Fig. 6). NICAP showed poor comparability with the arterial line during anesthesia induction. Moreover, NICAP showed poor ability to detect blood pressure change trends compared with the arterial line.

Fig. 2
figure 2

Bland-Altman analysis between NICAP and ART for SBP, DBP, and MAP

Fig. 3
figure 3

Scatter plots and linear regression results for SBP, DBP, and MAP

Fig. 4
figure 4

Polar plot analysis on trending ability of changes of systolic (A), diastolic (B), and mean arterial blood pressure (C) measured by NICAP against the arterial line

Fig. 5
figure 5

Bland-Altman analysis between NICAP and ART for SBP, DBP and MAP during anesthesia induction

Fig. 6
figure 6

Scatter plots and linear regression results for SBP, DBP, and MAP during anesthesia induction

Discussion

Our results demonstrated that blood pressure measured by NICAP showed poor correlation with arterial blood pressure in elderly patients either during the whole surgery or during anesthesia induction. Moreover, NICAP had poor comparability with an arterial line in the detection of blood pressure change trends. There were no major complications that occurred either from the arterial line or NICAP.

Recent studies focusing on non-invasive blood pressure monitoring devices suggest that continuous non-invasive beat-to-beat monitoring of blood pressure can effectively and instantaneously detect hypotension during surgeries compared with traditional cuff blood pressure monitoring that usually occurs 3-5 min intermittently [30,31,32,33]. However, these continuous non-invasive monitoring devices use different mechanisms and algorithms and showed great heterogeneity in the accuracy of detecting the actual blood pressure. Multiple studies concluded that these devices could not replace traditional cuff and invasive monitoring techniques [34, 35]. Similar to these studies, we also found that NICAP showed a poor correlation with the arterial blood pressure in elderly patients regardless of SBP, DBP, or MAP during the whole surgery. Several reasons might contribute to its inaccuracy. First, the NICAP system algorithm was formulated using intraoperative monitoring data from adults between 18 and 70. However, the mean age of the studied population was 72.29 years old, with the oldest of 87 years old. Aging is associated with many physiological changes in large and small arteries’ structural and functional properties, including arterial diameter, wall thickness, wall stiffness, and endothelial function [36, 37]. As a consequence of increased arterial stiffness, pulse wave velocity increases, and reflected waves return faster [38, 39]. This was a key element in the algorithm of NICAP. Second, comorbidities were generally common in elderly patients. In this study, hypertension was present in 20 (48.8%) and diabetes 37 (90.2%) out of 41 patients. Hypertension and diabetes could also cause many physiological alterations. For elderly patients with multiple comorbidities, the NICAP algorithm might need updates and iteration before clinical application. Third, the influence of involuntary body movements might be a major reason for the inaccuracy. The NICAP system requires two pulse wave readings from ear and toe clamps to correct blood pressure calculation. These two clamps are relatively far from each other in different body parts and thus could be easily affected due to the anesthesia and surgical settings in the operating room. Although muscle relaxants were administered, procedures like bag-valve ventilation and endotracheal intubation during induction could accidentally affect the ear clamp. Correspondingly, surgical manipulation could also affect the ear and toe clamps during the surgery, causing inaccuracy. Compared to other continuous blood pressure monitoring systems using artery contouring techniques, this incidence might be higher [24]. Because of this particular characteristic of NICAP, appropriate surgery type and sites might be a prerequisite for successful application.

Blood pressure readings from NICAP could easily be affected during induction, as aforementioned. Anesthesia induction is considered the most critical phase in surgery due to the most unstable hemodynamics, especially for elderly patients [10, 40]. Accurate detection of hemodynamic changes during induction could facilitate timely treatment from the anesthesiologists, thus avoiding catastrophic events. Although NICAP could be easily established in a timely manner, the inaccuracy limited its application in this high dependent situation. In our daily practice, the arterial line was seldomly established before induction due to pain stimulation while cuff blood pressure monitoring takes relatively longer per test. The application of NICAP during induction could have been an excellent alternative. However, further research, development, and calibration are still needed before it could be applied to elderly patients in situations like this.

NICAP also showed poor comparability with the arterial line in the detection of blood pressure change trends during the surgery. Worthy of note, in the present study, the concordance rate of above 90% was considered has good trending ability. However, this standard was adopted from studies focusing on cardiac output devices [27, 28]. To our knowledge, there was no consensus for good trending ability of blood pressure monitoring devices. Although NICAP has poor trending ability according to this standard, it is still considered by the authors as a good device for anesthesiologists to monitor the blood pressure change trend continuously. It could facilitate timely reactions of anesthesiologists to treat potential hyper- or hypotension episodes before it occurs. Another advantage we found in this study was that NICAP took significantly less time than the arterial line to establish accountable continuous blood pressure readings. We encountered one patient in the PACU that was complicated with postoperative bleeding in the study period. The patient was hypotensive and in a hemorrhagic shock. Traditional cuff blood pressure readings could not be obtained due to low circulating volume, and the arterial line was challenging to establish due to hypotension. In this situation, the NICAP acquired accountable continuous blood pressure readings in a short time. Although arterial line could not be replaced in many clinical cases, timely establishment of continuous blood pressure readings by NICAP could guide the resuscitation and rescue of critical patients giving that the trend of blood pressure changes could be accurately reflected by NICAP. This needs to be further studied in the future.

No complications such as hematoma, pressure injury was recorded in the study. The clamps are light-weighted, and the force applied to the ear and toe is relatively small. This might facilitate prolonged monitoring of non-invasive continuous blood pressure by NICAP in a non-high dependency situation for those high-risk patients. However, the mean surgical time in this study was less than 2 h and could not be interpreted as a prolonged period. Future studies focusing on prolonged monitoring and potential complications should be conducted before a conclusion is drawn.

There were several limitations of this study. Firstly, despite 6751 pairs of valid data analyzed, the sample size of this study was small. Future studies with more subjects are needed. Secondly, the NICAP has many restrictions, and these restrictions may affect its clinical application. Thus, the patients included in this study might not represent all the surgical types. The results of this study should be interpreted with caution. Finally, the mean surgical time of this study was 167.83 min which was less than 3 h. Longer duration might cause some complications such as pressure injury by ear and toe clamps of NICAP. Future studies are warranted to investigate these issues.

Conclusions

In summary, this study has demonstrated that NICAP has a poor correlation with the arterial lines in elderly patients either during the whole surgery or during anesthesia induction. Moreover, it showed poor comparability with the arterial line in the detection of blood pressure change trends. Our findings suggest that NICAP might not be sufficiently accurate to be applied clinically in elderly patients. More precise calibration and iteration of NICAP are needed for the elderly population.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Change history

Abbreviations

NICAP:

Non-Invasive Continuous Arterial Pressure system

CNAP:

Continuous Non-invasive Arterial Pressure

AAMI:

The Association for the Advancement of Medical Instrumentation

PWV:

Pulse wave velocity

SBP:

Systolic blood pressure

DBP:

Diastolic blood pressure

MAP:

Mean blood pressure

PACU:

Post-anesthesia care unit

OR:

Operating room

PEEP:

Positive end-expiratory pressure

EtCO2:

End-Tidal Carbon Dioxide

SD:

Standard deviations

IQR:

25th to 75th percentiles

LOAs:

Limit of agreements

References

  1. Mathis MR, Naik BI, Freundlich RE, Shanks AM, Heung M, Kim M, et al. Preoperative risk and the association between hypotension and postoperative acute kidney injury. Anesthesiology. 2020;132:461–75.

    Article  Google Scholar 

  2. Walsh M, Devereaux PJ, Garg AX, Kurz A, Turan A, Rodseth RN, et al. Relationship between intraoperative mean arterial pressure and clinical outcomes after noncardiac surgery: toward an empirical definition of hypotension. Anesthesiology. 2013;119:507–15.

    Article  Google Scholar 

  3. Mascha EJ, Yang D, Weiss S, Sessler DI. Intraoperative mean arterial pressure variability and 30-day mortality in patients having noncardiac surgery. Anesthesiology. 2015;123:79–91.

    Article  Google Scholar 

  4. Monk TG, Bronsert MR, Henderson WG, Mangione MP, Sum-Ping STJ, Bentt DR, et al. Association between intraoperative hypotension and hypertension and 30-day postoperative mortality in noncardiac surgery. Anesthesiology. 2015;123:307–19.

    Article  Google Scholar 

  5. Sun LY, Wijeysundera DN, Tait GA, Beattie WS. Association of intraoperative hypotension with acute kidney injury after elective noncardiac surgery. Anesthesiology. 2015;123:515–23.

    Article  Google Scholar 

  6. Waes JAR, Klei WA, Dn W, Wolfswinkel L, Tf L, Ws B. Association between intraoperative hypotension and myocardial injury after vascular surgery. Anesthesiology. 2016;124:35 44.

    PubMed  Google Scholar 

  7. Salmasi V, Maheshwari K, Yang D, Mascha EJ, Singh A, Sessler DI, et al. Relationship between intraoperative hypotension, defined by either reduction from baseline or absolute thresholds, and acute kidney and myocardial injury after noncardiac surgery: A retrospective cohort analysis. Anesthesiology. 2017;126:47–65.

    Article  Google Scholar 

  8. Abbott TEF, Pearse RM, Archbold RA, Ahmad T, Niebrzegowska E, Wragg A, et al. A prospective international multicentre cohort study of intraoperative heart rate and systolic blood pressure and myocardial injury after noncardiac surgery: Results of the VISION study. Anesth Analg. 2018;126:1936–45.

    Article  Google Scholar 

  9. Futier E, Lefrant J-Y, Guinot P-G, Godet T, Lorne E, Cuvillon P, et al. Effect of individualized vs standard blood pressure management strategies on postoperative organ dysfunction among high-risk patients undergoing major surgery: A randomized clinical trial. JAMA. 2017;318:1346–57.

    Article  Google Scholar 

  10. Südfeld S, Brechnitz S, Wagner JY, Reese PC, Pinnschmidt HO, Reuter DA, et al. Post-induction hypotension and early intraoperative hypotension associated with general anaesthesia. Br J Anaesth. 2017;119:57–64.

    Article  Google Scholar 

  11. Maheshwari K, Ahuja S, Khanna AK, Mao G, Perez-Protto S, Farag E, et al. Association between perioperative hypotension and delirium in postoperative critically ill patients: A retrospective cohort analysis. Anesth Analg. 2020;130:636–43.

    Article  Google Scholar 

  12. Wesselink EM, Kappen TH, Torn HM, Slooter AJC, van Klei WA. Intraoperative hypotension and the risk of postoperative adverse outcomes: a systematic review. Br J Anaesth. 2018;121:706–21.

    Article  CAS  Google Scholar 

  13. Silva A de. Anesthetic monitoring. In: Basics of anesthesia. Elsevier; 2011. p. 319–33.

  14. Bartels K, Esper SA, Thiele RH. Blood pressure monitoring for the anesthesiologist. Anesth Analg. 2016;122:1866–79.

    Article  CAS  Google Scholar 

  15. Tao G, Chen Y, Wen C, Bi M. Statistical analysis of blood pressure measurement errors by oscillometry during surgical operations. Blood Press Monit. 2011;16:285–90.

    Article  Google Scholar 

  16. Chen Y, Wen C, Tao G, Bi M. A new methodology of continuous and noninvasive blood pressure measurement by pulse wave velocity. In: 2010 11th International Conference on Control Automation Robotics & Vision. IEEE; 2010. doi:https://doi.org/10.1109/icarcv.2010.5707813.

  17. Chen Y, Wen C, Tang H, Xiuzhen Teng L. The relationship between different pulse wave velocity and systolic/diastolic pressure. In: 2008 3rd IEEE Conference on Industrial Electronics and Applications. IEEE; 2008. doi:https://doi.org/10.1109/iciea.2008.4582705.

  18. Chen Y, Wen C, Tao G, Bi M. Continuous and noninvasive measurement of systolic and diastolic blood pressure by one mathematical model with the same model parameters and two separate pulse wave velocities. Ann Biomed Eng. 2012;40:871–82.

    Article  Google Scholar 

  19. Akkermans J, Diepeveen M, Ganzevoort W, van Montfrans GA, Westerhof BE, Wolf H. Continuous non-invasive blood pressure monitoring, a validation study of Nexfin in a pregnant population. Hypertens Pregnancy. 2009;28:230–42.

    Article  CAS  Google Scholar 

  20. Tanioku T, Yoshida A, Aratani Y, Fujii K, Kawamata T. Validation of noninvasive continuous arterial pressure measurement by ClearSight SystemTM during induction of anesthesia for cardiovascular surgery. BMC Anesthesiol. 2020;20:176.

    Article  Google Scholar 

  21. Vos JJ, Poterman M, Mooyaart EAQ, Weening M, Struys MMRF, Scheeren TWL, et al. Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia. Br J Anaesth. 2014;113:67–74.

    Article  CAS  Google Scholar 

  22. Ameloot K, Van De Vijver K, Van Regenmortel N, De Laet I, Schoonheydt K, Dits H, et al. Validation study of Nexfin® continuous non-invasive blood pressure monitoring in critically ill adult patients. Minerva Anestesiol. 2014;80:1294–301.

    CAS  PubMed  Google Scholar 

  23. Schramm P, Tzanova I, Gööck T, Hagen F, Schmidtmann I, Engelhard K, et al. Noninvasive hemodynamic measurements during neurosurgical procedures in sitting position. J Neurosurg Anesthesiol. 2017;29:251–7.

    Article  Google Scholar 

  24. Weiss E, Gayat E, Dumans-Nizard V, Le Guen M, Fischler M. Use of the NexfinTM device to detect acute arterial pressure variations during anaesthesia induction. Br J Anaesth. 2014;113:52–60.

    Article  CAS  Google Scholar 

  25. Heusdens JF, Lof S, Pennekamp CWA, Specken-Welleweerd JC, de Borst GJ, van Klei WA, et al. Validation of non-invasive arterial pressure monitoring during carotid endarterectomy. Br J Anaesth. 2016;117:316–23.

    Article  CAS  Google Scholar 

  26. Jeleazcov C, Krajinovic L, Münster T, Birkholz T, Fried R, Schüttler J, et al. Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth. 2010;105:264–72.

    Article  CAS  Google Scholar 

  27. Critchley LA, Lee A, Ho AM-H. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.

    Article  Google Scholar 

  28. Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25:536–46.

    Article  Google Scholar 

  29. Chen G, Chung E, Meng L, Alexander B, Vu T, Rinehart J, et al. Impact of non invasive and beat-to-beat arterial pressure monitoring on intraoperative hemodynamic management. J Clin Monit Comput. 2012;26:133–40.

    Article  CAS  Google Scholar 

  30. Ilies C, Kiskalt H, Siedenhans D, Meybohm P, Steinfath M, Bein B, et al. Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement. Br J Anaesth. 2012;109:413–9.

    Article  CAS  Google Scholar 

  31. Meidert AS, Nold JS, Hornung R, Paulus AC, Zwißler B, Czerner S. The impact of continuous non-invasive arterial blood pressure monitoring on blood pressure stability during general anaesthesia in orthopaedic patients. Eur J Anaesthesiol. 2017;34:716–22.

    Article  Google Scholar 

  32. Benes J, Simanova A, Tovarnicka T, Sevcikova S, Kletecka J, Zatloukal J, et al. Continuous non-invasive monitoring improves blood pressure stability in upright position: randomized controlled trial. J Clin Monit Comput. 2015;29:11–7.

    Article  Google Scholar 

  33. Stenglova A, Benes J. Continuous non-invasive arterial pressure assessment during surgery to improve outcome. Front Med (Lausanne). 2017;4. doi:https://doi.org/10.3389/fmed.2017.00202.

  34. Harju J, Vehkaoja A, Kumpulainen P, Campadello S, Lindroos V, Yli-Hankala A, et al. Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement. J Clin Monit Comput. 2018;32:13–22.

    Article  Google Scholar 

  35. Hohn A, Defosse JM, Becker S, Steffen C, Wappler F, Sakka SG. Non-invasive continuous arterial pressure monitoring with Nexfin does not sufficiently replace invasive measurements in critically ill patients. Br J Anaesth. 2013;111:178–84.

    Article  CAS  Google Scholar 

  36. Najjar SS, Scuteri A, Lakatta EG. Arterial aging: is it an immutable cardiovascular risk factor? Hypertension. 2005;46:454–62.

    Article  CAS  Google Scholar 

  37. Oh YS. Arterial stiffness and hypertension. Clin Hypertens. 2018;24:17.

    Article  Google Scholar 

  38. O’Rourke MF, Adji A. An updated clinical primer on large artery mechanics: implications of pulse waveform analysis and arterial tonometry. Curr Opin Cardiol. 2005;20:275–81.

    Article  Google Scholar 

  39. Lee J-H, Kim K-I, Cho M-C. Current status and therapeutic considerations of hypertension in the elderly. Korean J Intern Med. 2019;34:687–95.

    Article  CAS  Google Scholar 

  40. Chen EY, Michel G, Zhou B, Dai F, Akhtar S, Schonberger RB. An analysis of anesthesia induction dosing in female older adults. Drugs Aging. 2020;37:435–46.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work is financially supported by the National Key R&D Program of China, Project Number: 2018YFC2001804-5 and the 1.3.5 project for disciplines of excellence, West China Hospital, Sichuan University, Grant Number: 20HXFH030 and 20HXJS019.

Author information

Authors and Affiliations

Authors

Contributions

ZX helped with substantial contributions to the conception and design of the work; the analysis and interpretation of the data and wrote the manuscript. HYC helped with implementing the study. HYZ helped with the analysis, interpretation of the data for the work, and editing the manuscript. XHS, JR and HXS helped with the acquisition of the data for the work. CC and GC helped edit and revise the manuscript critically for important intellectual content and final approval of the version to be published. The authors read and approved the final manuscript.

Corresponding authors

Correspondence to Chan Chen or Guo Chen.

Ethics declarations

Ethics approval and consent to participate

The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013). This study was approved by the Ethics Committee of Clinical and Biomedical Trials, West China Hospital of Sichuan University (No. 2019-136). Written informed consent was obtained from all participants.

Consent for publication

Not applicable.

Competinginterests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Xu, Z., Chen, H., Zhou, H. et al. Comparison of noninvasive continuous arterial blood pressure measured by NICAP with arterial line in elderly patients. BMC Geriatr 22, 108 (2022). https://doi.org/10.1186/s12877-022-02803-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12877-022-02803-3

Keywords